LSA-01 Split Style Laboratory Syringe Pump
LSA-01 is a single-channel microsyringe pump with five modes: infusion, withdrawal, infusion/withdrawal, withdrawal/infusion, and continuous cycling. It features a split structure design, with the execution unit equipped with a mounting bracket for seamless integration with microscopic operating equipment, stereotaxic instruments, and other fixtures. It's designed for microliter syringes (5µL-1000µL), offering precise control (1μm/min to 150mm/min). It features a user-friendly color touch LCD screen for parameter adjustments and has built-in RS485 Modbus support for external control devices.
Functions and Features
- Color touch LCD screen and physical function keys
- Multiple working modes: infusion, withdrawal, infusion-then-withdrawal, withdrawal-then-infusion, and continuous cycling
- Split-style design for easy installation and operation
- Compatibility with various syringe types, ensuring high precision
- RS485 MODBUS interface for control
- Durable full metal housing.
Specifications
- Syringe Size: 5μL-1000μL
- Number of Channels: 1
- Max Stroke Range: 70 mm
- Stroke Resolution: 0.198 um/step
- Linear Velocity Range: 0.635 um/min to 95.25 um/min
- Linear Velocity Regulation Resolution: 0.635 um/min
- Linear Travel Accuracy: ±0.5% (at 30% stroke)
- Maximum Linear Thrust: > 20n
- Thrust Adjustment: 1-100% adjustable
- Syringe Selection: Built-in syringe options or custom input
- Flow Correction: Calibration for precise volume control
- Display: 4.3-inch color LCD with 480x272 resolution and 65K colors
- Displayed Parameters: Includes transfusion volume, liquid quantity, flow rate, direction, syringe specifications, and more
- Animation Display for Operational Status
- Operation Modes: Touch screen and physical buttons
- Power-off Memory: Retains saved parameters
- Additional Features: Pause and stop functions, warning tone, parameter lock
- Fast Forward and Retreat Options
- Indicator and State Signal Outputs: One for stop state, another for direction state
- Control Signal Input: Start/stop and direction control
- Communication Interface: RS485, MODBUS
- Power Supply: AC100-240V, 50/60Hz
- Operating Temperature: 5-40°C (41-104°F)
- Operating Environment Relative Humidity: < 80%
- Controller Dimensions: 245x195x133 mm (9.6x7.7x5.2 inches)
- Drive Dimensions: 70x44x194 mm (2.8x1.7x7.6 inches)
- Weight: 3.2 kg (7.0 lbs)
Reference Flow Rate
Syringe size | Syringe ID(mm) | Min flow rate (µl/min) | Max flow rate(mL/min) |
5µl | 0.343 | 0.058 | 0.009 |
10µl | 0.485 | 0.117 | 0.017 |
25µl | 0.728 | 0.264 | 0.039 |
50µl | 1.03 | 0.529 | 0.079 |
100µl | 1.457 | 1.058 | 0.158 |
250µl | 2.303 | 2.645 | 0.396 |
500µl | 3.257 | 5.290 | 0.793 |
1000µl | 4.606 | 10.580 | 1.587 |