WMD15 Low-pulsation Pump Head

WMD15 Low-pulsation Pump Head

  • €595,00
    Precio unitario por 
    (excl. VAT)
Los gastos de envío se calculan en la pantalla de pagos.


The WMD15 series low-pulsation pump head offers a flow rate range of 0.006–1740 mL/min and is compatible with tubing sizes 13#, 14#, 16#, 17#, 19#, and 25#. Featuring proprietary low-pulsation technology, it enables smoother fluid transfer with reduced pulsation and improved flow continuity, making it especially suitable for high-stability applications such as dispensing, coating, and chemical synthesis. Constructed from aluminum alloy and SS304 stainless steel, the pump head is robust and durable, supporting reliable long-term continuous operation at high speeds.

Application fields: laboratories, new energy, environmental protection, and chemical industries.

Typical applications: online COD water quality analyzers, automatic titrators, and form-fill-seal machines.

Specifications

  • Flow Rate Range: 0.006–1740 mL/min
  • Speed Range: 0.1–600 rpm
  • Number of Rollers: 3
  • Occlusion (Gap Adjustment): Fixed, optimized for dedicated tubing
  • Tubing Compatibility:
    • WMD15-14: 13#, 14#
    • WMD15-16: 16#, 19#
    • WMD15-17: 17#, 25#
  • Operating Temperature: 0°C to 40°C
  • Pump Head Materials:
    • Housing: Aluminum alloy
    • Rollers and Roller Shafts: SS304 stainless steel

Flow Rate Parameters

Tubing Materials: Silicone, Pharmed, Viton, Chemical-resistant tubing, A-60-G/F
Wall Thickness: 1.6 mm

Tubing Size

ID (mm)

Flow (mL/rev)

Max Flow (mL/min)

max. 600 rpm continuous

13# 0.8 0.06 36
14# 1.6 0.23 130
19# 2.4 0.48 286
16# 3.1 0.83 498
25# 4.8 1.67 1000
17# 6.4 2.90 1740


Flow Stability Comparison

  • A: Standard peristaltic pump head, 14# silicone tubing, speed 20 rpm
  • B: WMD15 pump head, 14# silicone tubing, speed 20 rpm

Tubing Life Comparison

Note: The data provided above were obtained through controlled testing using pure water at standard temperature and pressure and are intended for reference purposes only. Please be advised that actual service life may vary based on specific operating conditions, including system pressure, temperature, fluid media characteristics, tubing batch variations, and wall thickness.

Dimension